Autor(es) Stuart D.G., Greenier T.J., Rumery R.A., Eagleson J.M.
Título Survey, use, and performance of biological safety cabinets
Información bibliográfica Apr. 1982, Vol.43, No.4, p.265-270. Illus. 18 ref.
Resumen The types of cabinet available, how they are designed to function, their proper use and their expected performance are discussed. A class II Type A biological safety cabinet was evaluated by challenge with Bacillus subtilis spores at line voltages of 85-130V in order to determine performance during simulated brownout and power surge conditions. With a 20.3cm high work opening the cabinet passed both personnel and product protection tests at all voltages. With a 25.4cm high work opening the cabinet passed the same tests with the exception of the personnel protection test at 85V. The motor-blower performance characteristics were adequate for this cabinet to perform its safety function over a considerable range of line voltages.
Descriptores (primarios) campanas de aspiración; riesgos biológicos; trabajos de laboratorio
Descriptores (secundarios) aspiración localizada; microrganismos; valoración del material
Tipo de documento D - Artículos periódicos