ID (ISN) del documento | 38566 |
Número CIS |
82-1634 |
ISSN - Título de la serie |
0002-8894 - American Industrial Hygiene Association Journal |
Año |
1982 |
Número de serie |
|
Autor(es) |
Stuart D.G., Greenier T.J., Rumery R.A., Eagleson J.M. |
Título |
Survey, use, and performance of biological safety cabinets |
Información bibliográfica |
Apr. 1982, Vol.43, No.4, p.265-270. Illus. 18 ref. |
Resumen |
The types of cabinet available, how they are designed to function, their proper use and their expected performance are discussed. A class II Type A biological safety cabinet was evaluated by challenge with Bacillus subtilis spores at line voltages of 85-130V in order to determine performance during simulated brownout and power surge conditions. With a 20.3cm high work opening the cabinet passed both personnel and product protection tests at all voltages. With a 25.4cm high work opening the cabinet passed the same tests with the exception of the personnel protection test at 85V. The motor-blower performance characteristics were adequate for this cabinet to perform its safety function over a considerable range of line voltages. |
Descriptores (primarios) |
campanas de aspiración; riesgos biológicos; trabajos de laboratorio |
Descriptores (secundarios) |
aspiración localizada; microrganismos; valoración del material |
Tipo de documento |
D - Artículos periódicos |
Tema(s) |
Sustancias tóxicas y peligrosas
|
Broad subject area(s) |
|
Navegación por categoria(s) |
Ventilation Biological hazards
|