ILO Home
Go to the home page
Site map | Contact us Français | Español

CISDOC database

Document ID (ISN)55954
CIS number 91-934
ISBN(s) 91-7045-038-2
Year 1989
Convention or series no. Arbete och hälsa 1989:26
Author(s) Lindahl R., Levin J.O., Andersson K.
Title Validation of a diffusive sampler for reactive compounds
Original title Utvärdering av en diffusionsprovtagare för reaktiva ämnen [in Swedish]
Bibliographic information Arbetsmiljöinstitutet, Förlagstjänst, 171 84 Solna, Sweden, 1989. 21p. Illus. 8 ref.
Abstract A commercially available diffusive device for reactive compounds has been validated for sampling of formaldehyde in laboratory and field trials. The sampler consists of a filter impregnated with 2,4-dinitrophenylhydrazine, phosphoric acid and glycerine mounted in a polypropylene housing. The uptake rate was 25.2mL/min, with a relative standard deviation of 7%. The uptake rate was independent of sampling time, relative humidity and concentration. The uptake rate was only slightly affected when air velocities ranged from 0.05 to 1m/s, and field trials showed that the sampler could be used at very low wind velocities (<0.02m/s). Sensitivity was 2µg/m3 in a 24-hour sample, making the sampler useful for indoor air monitoring of low formaldehyde levels. The sampler can be stored at least 3 months at -18°C prior to sampling.
Descriptors (primary) formaldehyde; air sampling; passive dosimeters
Descriptors (secondary) report; equipment testing; sampling instruments
Document type E - Books, reports, proceedings
Country / State or ProvinceSweden
Subject(s) Toxic and dangerous substances
Broad subject area(s) Chemical safety
Browse category(ies) Aldehydes