ILO Home
Go to the home page
Site map | Contact us Français | Español

CISDOC database

Document ID (ISN)50576
CIS number 88-1952
ISSN - Serial title 0002-8894 - American Industrial Hygiene Association Journal
Year 1987
Convention or series no.
Author(s) Naus C.J., McAvoy D., Broder I., Smith J.W.
Title A non-toxic diffusional dosimeter for sulfur dioxide
Bibliographic information Dec. 1987, Vol.48, No.12, p.1001-1003. Illus. 4 ref.
Abstract The method combined the use of a diffusional dosimeter and a relatively non-toxic buffered formaldehyde absorbing solution. The dosimeter gave reproducible results using two lots of single-backed dimethylsilicone membranes; sampling times from 1 to 7hr; sampling temperatures from 25°C to 38°C; and at face velocities from 0-600ft/min (0-3m/s). Concentrations as low as 0.05ppm were determined over a maximum sampling time of 7hr, yielding a sensitivity better than 0.35ppm.hr. Fluctuating concentrations from 0.5, 1.0 or 2.0ppm background to 10.0ppm for 1/2hr resulted in no significant error.
Descriptors (primary) dosimetry; sulfur dioxide; determination in air; passive dosimeters
Descriptors (secondary) evaluation of equipment; description of equipment
Document type D - Periodical articles
Subject(s) Toxic and dangerous substances
Broad subject area(s) Chemical safety
Browse category(ies) Inorganic sulfur compounds
Inorganic substances