Abstract The types of cabinet available, how they are designed to function, their proper use and their expected performance are discussed. A class II Type A biological safety cabinet was evaluated by challenge with Bacillus subtilis spores at line voltages of 85-130V in order to determine performance during simulated brownout and power surge conditions. With a 20.3cm high work opening the cabinet passed both personnel and product protection tests at all voltages. With a 25.4cm high work opening the cabinet passed the same tests with the exception of the personnel protection test at 85V. The motor-blower performance characteristics were adequate for this cabinet to perform its safety function over a considerable range of line voltages.