Abstract |
The disturbing effects of these factors on detector tubes are described. Humidity has no influence on tubes of modern design. The effects of air pressure and temperature on the indicated values can be, in general, corrected by simple arithmetic. When temperatures are <0°C or >40°C, techniques must be used to bring the tube temperature into these limits during the measurements. Examples of the specific cases studied. French translation may be obtained from INRS, 30 rue Olivier-Noyer, 75680 Paris Cedex 14, France. |
Descriptors (primary) |
hyperbaric pressure; humidity; extreme temperatures; detector tubes; toxic atmosphere detection |
Descriptors (secondary) |
accuracy; description of technique; translation; mathematical analysis |