ILO Home
Go to the home page
Site map | Contact us Français | Español

CISDOC database

Document ID (ISN)27620
CIS number 77-1341
Year 1977
Convention or series no.
Author(s) Skare I.
Title Evaluation of indicator tubes - Part IV: Carbon dioxide
Original title Utvärdering av vissa analysampullers tillförlitlighet - IV. Koldioxid [in Swedish]
Bibliographic information Undersökningsrapport 1977:1, Arbetarskyddsstyrelsen, Fack, 100 26 Stockholm, Sweden, 1977. 19p. 3 ref. Price: Swe-cr.5.00.
Abstract Seven tubes (Auer 817; Bacharach 19-199; Dräger CH 235; Gastec 2LL, 2L, 2H; MSA 85976) were tested for accuracy and effects of temperature and humidity. The accuracy was acceptable for all types, and there were no interfering substances at the high carbon dioxide concentrations. At extremely high humidities, all tubes gave positive errors. All the tubes were acceptable except the Gastec 2LL (because of the measuring range). English summary.
Descriptors (primary) carbon dioxide; detector tubes; determination in air
Descriptors (secondary) evaluation of equipment; report; Sweden
Document type E - Books, reports, proceedings
Subject(s) Toxic and dangerous substances
Broad subject area(s)
Browse category(ies) Carbon dioxide
Inorganic substances