Title |
A comparison of controlled negative pressure and aerosol quantitative respirator fit test systems by using fixed leaks |
Bibliographic information |
June 1991, Vol.52, No.6, p.249-251. Illus. 3 ref. |
Abstract |
An automated version of a new method for quantitative respirator fit testing by controlled negative pressure was compared with a computerised aerosol fit test system. The controlled negative pressure technique eliminates many of the problems associated with aerosol and pressure decay fit test methods. A series of fixed leaks was used to compare the leak measurement capabilities of the controlled negative pressure system against a standard computerised aerosol fit test system. Negative pressure and aerosol fit factors determined for a series of fixed leaks through hypodermic needles were highly correlated with each other (r=0.998) and with the cross-sectional areas of the leak needles (r>0.995). |
Descriptors (primary) |
negative pressure application; respirators; leakage testing; equipment testing; facepieces |
Descriptors (secondary) |
computer applications; automation |